高低溫濕熱試驗箱

    1.在高溫(wen)試(shi)驗中,如溫(wen)度(du)變(bian)化(hua)達不到試(shi)驗溫(wen)度(du)值時,可以檢查電(dian)器系統,逐(zhu)一排除故(gu)障(zhang)。如溫(wen)度(du)升得(de)很慢(man),就要查看風循(xun)環系統,看一下風循(xun)環的調節擋(dang)板是否(fou)(fou)開(kai)啟正(zheng)常(chang),反(fan)之,就檢查風循(xun)環的電(dian)機(ji)運轉是否(fou)(fou)正(zheng)常(chang)。如溫(wen)度(du)過沖厲害那么(me)就需要整(zheng)定PID的設(she)置參數(shu)。如果溫(wen)度(du)直接上升,過溫(wen)保護,那么(me),控(kong)制(zhi)器出(chu)故(gu)  障(zhang),須更換控(kong)制(zhi)儀表。

    2.低溫(wen)達不(bu)到試(shi)驗的(de)指標(biao),那(nei)你就要(yao)(yao)觀(guan)察溫(wen)度(du)的(de)變化,是(shi)(shi)溫(wen)度(du)降(jiang)的(de)很慢,還是(shi)(shi)溫(wen)度(du)到一定(ding)(ding)值后溫(wen)度(du)有回(hui)升的(de)趨勢,前者就要(yao)(yao)檢查(cha)一下,做低溫(wen)試(shi)驗前是(shi)(shi)否將工(gong)作(zuo)室烘干,使工(gong)作(zuo)室保持(chi)干燥后再將試(shi)驗樣(yang)品(pin)放入工(gong)作(zuo)室內再做試(shi)驗,工(gong)作(zuo)室內的(de)試(shi)驗樣(yang)品(pin)是(shi)(shi)否放置(zhi)的(de)過多,使工(gong)作(zuo)室內的(de)風不(bu)能充分(fen)循環,在排除上(shang)述(shu)原因后,就要(yao)(yao)考(kao)慮是(shi)(shi)否是(shi)(shi)制冷(leng)系統中(zhong)的(de)故障了,這樣(yang)就要(yao)(yao)請廠家(jia)的(de)專業人員(yuan)進行檢修。后者的(de)現象是(shi)(shi)設備的(de)使用環境不(bu)好(hao)所致,設備放置(zhi)的(de)環境溫(wen)度(du),放置(zhi)的(de)位置(zhi)(箱體后與墻的(de)距離(li))要(yao)(yao)滿足要(yao)(yao)求(qiu)(在設備操作(zuo)使用說明中(zhong)都(dou)有規定(ding)(ding))。

    3.在做濕(shi)熱試驗(yan)中(zhong),出現(xian)實際(ji)濕(shi)度(du)(du)會達(da)到100%或(huo)者實際(ji)濕(shi)度(du)(du)與目標(biao)濕(shi)度(du)(du)相差很(hen)大,數值低得很(hen)多(duo),前者的(de)(de)(de)(de)現(xian)象:可(ke)能(neng)是(shi)濕(shi)球(qiu)上(shang)的(de)(de)(de)(de)紗(sha)布(bu)(bu)干燥引起,那就要(yao)檢查濕(shi)球(qiu)傳感器的(de)(de)(de)(de)水(shui)(shui)(shui)(shui)槽(cao)中(zhong)是(shi)否缺水(shui)(shui)(shui)(shui),水(shui)(shui)(shui)(shui)槽(cao)中(zhong)的(de)(de)(de)(de)水(shui)(shui)(shui)(shui)位(wei)(wei)(wei)(wei)是(shi)由一(yi)(yi)水(shui)(shui)(shui)(shui)位(wei)(wei)(wei)(wei)控制(zhi)(zhi)器自動(dong)控制(zhi)(zhi)的(de)(de)(de)(de),查水(shui)(shui)(shui)(shui)位(wei)(wei)(wei)(wei)控制(zhi)(zhi)器供(gong)水(shui)(shui)(shui)(shui)系(xi)統是(shi)否供(gong)水(shui)(shui)(shui)(shui)正(zheng)(zheng)常,水(shui)(shui)(shui)(shui)位(wei)(wei)(wei)(wei)控制(zhi)(zhi)器工(gong)作(zuo)是(shi)否正(zheng)(zheng)常。另一(yi)(yi)種可(ke)能(neng)就是(shi)濕(shi)球(qiu)紗(sha)布(bu)(bu)因使用時(shi)間(jian)長,或(huo)供(gong)水(shui)(shui)(shui)(shui)水(shui)(shui)(shui)(shui)質(zhi)純凈(jing)度(du)(du)的(de)(de)(de)(de)原因,會使紗(sha)布(bu)(bu)變硬,使紗(sha)布(bu)(bu)無法吸收水(shui)(shui)(shui)(shui)份而(er)干燥,只要(yao)更換或(huo)清洗紗(sha)布(bu)(bu)即(ji)可(ke)排除以(yi)上(shang)現(xian)象。后者的(de)(de)(de)(de)現(xian)象主要(yao)是(shi)加濕(shi)系(xi)統不工(gong)作(zuo),查看加濕(shi)系(xi)統的(de)(de)(de)(de)供(gong)水(shui)(shui)(shui)(shui)系(xi)統,供(gong)水(shui)(shui)(shui)(shui)系(xi)統內(nei)是(shi)否有一(yi)(yi)定的(de)(de)(de)(de)水(shui)(shui)(shui)(shui)量,控制(zhi)(zhi)加濕(shi)鍋爐水(shui)(shui)(shui)(shui)位(wei)(wei)(wei)(wei)的(de)(de)(de)(de)水(shui)(shui)(shui)(shui)位(wei)(wei)(wei)(wei)控制(zhi)(zhi)是(shi)否正(zheng)(zheng)常,加濕(shi)鍋爐內(nei)的(de)(de)(de)(de)水(shui)(shui)(shui)(shui)位(wei)(wei)(wei)(wei)是(shi)否正(zheng)(zheng)常。如(ru)以(yi)上(shang)一(yi)(yi)切(qie)都正(zheng)(zheng)常,那就要(yao)檢查電器控制(zhi)(zhi)系(xi)統,這要(yao)請專業維修人員(yuan)進行檢修。

    4.設(she)(she)備在(zai)試(shi)驗運(yun)行過(guo)程中突然出(chu)(chu)現(xian)故(gu)(gu)障(zhang)時,控制(zhi)儀表上(shang)出(chu)(chu)現(xian)對應的(de)故(gu)(gu)障(zhang)顯示提(ti)示并有聲訊(xun)報警提(ti)示。操作(zuo)人員(yuan)可以(yi)對照設(she)(she)備的(de)操作(zuo)使用中的(de)故(gu)(gu)障(zhang)排(pai)除一(yi)章中快(kuai)(kuai)速檢查出(chu)(chu)屬于哪(na)一(yi)類(lei)故(gu)(gu)障(zhang),即可請(qing)專業(ye)人員(yuan)快(kuai)(kuai)速排(pai)除故(gu)(gu)障(zhang),以(yi)確保(bao)試(shi)驗的(de)正常(chang)進行。其(qi)它環境(jing)試(shi)驗設(she)(she)備在(zai)使用中還會(hui)有其(qi)它的(de)現(xian)象,那就要具(ju)體(ti)現(xian)象,具(ju)體(ti)分析和排(pai)除。

    環境(jing)試驗(yan)設備還要定期(qi)進行(xing)維(wei)護保養,制冷(leng)系統的冷(leng)凝器(qi)定期(qi)清理,對于(yu)活動部件應按說明書(shu)加(jia)油潤滑,電(dian)器(qi)控(kong)制系統定期(qi)維(wei)護檢查等等,這(zhe)些工作是(shi)*的。